Uvandlakanyo lokusebenza phakathi kweSF-8200 kunye neStago Compact Max3


Umbhali: Ophumeleleyo   

微信图片_20211012132116

I-artical yapapashwa kwi-Clin.Lab.nguOguzhan Zengi, uSuat H. Kucuk.

Yintoni iClin.Lab.?

ILebhu yoNyango yijenali yelizwe jikelele ephononongwe ntanganye equka yonke imiba yonyango lwaselabhoratri kunye neyeza lotofelo-gazi.Ukongeza kwizihloko zonyango lotofelo-gazi iLabhoratri yeClinikhi imele ukuhanjiswa malunga nokufakelwa kwezicubu kunye ne-hematopoietic, iselula kunye nonyango lwemfuza.Ijenali ipapasha amanqaku okuqala, amanqaku okuphonononga, iipowusta, iingxelo ezimfutshane, izifundo zeemeko kunye neeleta kumhleli ojongene ne-1) imvelaphi yesayensi, ukuphunyezwa kunye nokubaluleka kokuxilongwa kweendlela zebhubhoratri eziqeshwe kwizibhedlele, iibhanki zegazi kunye neeofisi zoogqirha kunye ne-2) imiba yenzululwazi, yolawulo kunye neklinikhi yeyeza lotofelo-gazi kunye ne-3) ukongeza kwizihloko zonyango lotofelo-gazi i-Clinical Laboratory imele ukungeniswa malunga nokufakelwa kwezicubu kunye ne-hematopoietic, iselula kunye nonyango lwemfuza.

 

ilebhu yekliniki

Bajonge ukwenza uphononongo lokuthelekisa ukusebenza phakathi kwe-Succeeder SF-8200 kunye neStago Compact Max3 kuba

abahlalutyi be-coagulation abazenzekela ngokupheleleyo baye baba lelona candelo libalulekileyo kwiilabhoratri zeklinikhi.

Iindlela: Iimvavanyo ze-coagulation zesiqhelo zavavanywa, ezizezona zi-odolwe kakhulu kwiilabhoratri ezifana ne-PT, i-APTT, kunye ne-fibrinogen.

Iziphumo: Ii-coefficients of variation zihlolwe kwi-intra and inter-assay precision analysiss zingaphantsi kwe-5% ngokumele iiparameters ezivavanyiweyo.Ukuthelekisa phakathi kwabahlalutyi kubonise iziphumo ezilungileyo.Iziphumo ezifunyenwe yi-SF-8200 zibonise ukuthelekiswa okuphezulu kakhulu kwi-analyzers esetyenzisiweyo, kunye ne-coefficients yokulungelelanisa ukusuka kwi-0.953 ukuya kwi-0.976.Kuseto lwethu lwesiqhelo lwaselabhoratri, i-SF-8200 ifikelele kwisampulu yokuphumelela iimvavanyo ezingama-360 ngeyure.Akukho mpembelelo ibonakalayo kwiimvavanyo efunyenweyo kumanqanaba aphezulu e-hemoglobin, i-bilirubin, okanye i-triglycerides.

Isiphelo: Ukuqukumbela, i-SF-8200 yayiyi-analyzer echanekileyo, echanekileyo, kunye nethembekileyo yokuhlaziya i-coagulation kuvavanyo lwesiqhelo.Ngokokufunda kwethu, iziphumo zibonise ukusebenza kakuhle kobugcisa kunye nokuhlalutya.