SA-6000

I-Automated Blood Rheology Analyzer

1. Yenzelwe iLebhu yenqanaba eliNcinci-Phakathi.
2. Indlela yepleyiti yeCone ejikelezayo.
3. Ummakishi osemgangathweni ongengowaseNewtonian uphumelele iSatifikethi seSizwe saseTshayina.
4. Izilawuli zokuqala ezingezizo zeNewtonian, Izinto ezisetyenziswayo kunye nesicelo zenza isisombululo esipheleleyo.


Iinkcukacha zeMveliso

Intshayelelo yomhlalutyi

I-SA-6000 i-automated blood rheology analyser ithatha indlela yokulinganisa i-cone / plate.Imveliso ibeka uxinzelelo olulawulwayo kulwelo oluza kulinganiswa ngenjini ye-inertial torque ephantsi.I-shaft ye-drive igcinwa kwindawo ephakathi nge-low resistant magnetic levitation bearing, ehambisa uxinzelelo olunyanzeliswayo kulwelo oluza kulinganiswa kwaye intloko yayo yokulinganisa luhlobo lwe-cone-plate.Yonke i-mensuration ilawulwa ngokuzenzekelayo yikhompyutha.Umlinganiselo wokucheba unokusetwa ngokungacwangciswanga kuluhlu lwe-(1~200) s-1, kwaye unokulandela umkhondo we-curve ene-dimensional curve yereyithi yokucheba kunye ne-viscosity ngexesha lokwenyani.Umgaqo wokulinganisa uzotywe kwi-Newton Viscidity Theorem.

Inkcazo yobuGcisa

Umzekelo SA-6000
Umgaqo Indlela yokujikeleza
Indlela Indlela ye-cone plate
Ingqokelela yemiqondiso Itekhnoloji yokwahlulwa kwe-raster ephezulu echanekileyo
Imo yokusebenza /
Umsebenzi /
Ukuchaneka ≤±1%
CV CV≤1%
Izinga lokucheba (1-200)s-1
Viscosity (0 - 60) mPa.s
Phungula ucinzelelo (0-12000) mPa
Umthamo wesampuli ≤800ul
Inkqubo Ingxubevange ye-Titanium, i-jewel bear
Indawo yesampuli Indawo yesampulu engama-60 enerakhi enye
Uvavanyo lwetshaneli 1
Inkqubo yolwelo Impompo ye-peristaltic ecudiweyo kabini, Phanda ngesivamvo solwelo kunye nomsebenzi wokwahlula okuzenzekelayo kweplasma
Ujongano RS-232/485/USB
Ubushushu 37℃±0.1℃
Ulawulo Itshathi yolawulo ye-LJ enokugcina, umbuzo, umsebenzi wokushicilela;
Ulawulo lokuqala lwe-Non-Newtonian fluid kunye nesiqinisekiso se-SFDA.
Ulungelelwaniso I-Newtonian fluid ilinganiswe ngolwelo lwesizwe oluphambili lwe-viscosity;
Ulwelo olungelulo olwaseNewtonian luphumelele isatifikethi somgangatho wesizwe esimakishayo yi-AQSIQ yaseTshayina.
Ingxelo Vula

Inkqubo yokulinganisa

Kukho umsebenzi wokulinganisa kwisoftware yovavanyo lwesixhobo.I-standard viscosity liquid elungiselelwe yiZiko leSizwe loPhando lweMathiriyeli yamkelwa.

1. Kufuneka nini ulungelelwaniso:

1.1 Isixhobo sifakwe ekuqaleni.

1.2 Isixhobo siyasuswa, kutshintshwe isistim yekhompyutha okanye i-viscosity meter okanye kufakwe ezinye endaweni yazo.

1.3 Emva kokusebenzisa isixhobo ixesha elithile, kufumaniseka ukuba ixabiso elilinganisiweyo lesixhobo linokuphambuka okucacileyo.

☆ Qaphela: Ngaphambi kokuba isixhobo silinganiswe, indawo ethe tye yentshukumo yovavanyo kufuneka ihlengahlengiswe: beka imitha yenqanaba kwiqonga lentshukumo lovavanyo, kwaye ujike iqhosha lohlengahlengiso phantsi kwesixhobo ukuze ubeke amaqamza kwisangqa esincinci. imitha yomgangatho.

2. Ukulinganisa u-zero:

Ngaphandle kokongeza naluphi na ulwelo kwichibi lolwelo lovavanyo, nqakraza iqhosha elithi "Yongeza iSampuli eMgangatho" kwi [Calibration Interface], kuvela "input dialog box", ngenisa ixabiso le-viscosity: 0, cofa "Kulungile", kwaye isixhobo siya kuqalisa. uvavanyo lokulinganisa amanqaku;Inkqubo inceda ukugcina isiphumo solungelelwaniso lwe-zero.

3. Ulungelelwaniso lwe-viscosity fluid esemgangathweni:

3.1 Sebenzisa i-pipette ukongeza i-0.8ml yolwelo lwe-viscosity esemgangathweni kwichibi lolwelo lovavanyo, klikha iqhosha elithi "Add Standard Sample" kwi- [calibration interface], kwaye kuvela i-"input dialog box" , ngenisa ulwelo lwe-viscosity oluqhelekileyo oludityanisiweyo. test liquid pool Viscosity value, cofa iqhosha elithi "OK", kwaye isixhobo siya kuqalisa uvavanyo oluqhelekileyo lwe-viscosity fluid calibration;

3.2 Emva kokuba uvavanyo lolungelelwaniso luphelile, ijika eliluhlaza liya kuboniswa kwi-coordinate yeshear rate-viscosity coordinate;

3.3 Bonisa i-viscosity kunye neeparameters ze-viscosity fluid ezihambelana nazo zonke ii-curve zokulinganisa kwibhokisi "yesampula eqhelekileyo"

4. Cima ijika yokulinganisa

4.1 Kwibhokisi ethi "standard sample list", sebenzisa imouse ukukhetha iqela ledatha ethe tyaba.Ngeli xesha, idatha igutyungelwe ngumbala ohlaza okwesibhakabhaka, kwaye ijika elihambelanayo kwi-shear rate-viscosity coordinate ijika ibe yellow.Cofa iqhosha elithi "delete standard sample", Emva koko ijika lokulinganisa liyanyamalala kwii-coordinates, kwaye inombolo ehambelanayo kwibhokisi ethi "standard sample sample" iyanyamalala;

4.2 Gcina ubuncinci ijika le-calibration ye-zero point, enye ye-viscosity ephezulu (malunga ne-27.0mPa•s) kunye ne-viscosity ephantsi (malunga ne-7.0mPa•s) ukuqinisekisa uvavanyo oluqhelekileyo lwesixhobo.

☆ Qaphela: Nceda ungenzi imisebenzi yokulinganisa ngaphandle kokugunyaziswa, ukuze ungabangeli ukudideka kwiiparamitha zangaphakathi zenkqubo yesixhobo kwaye uchaphazele ukuchaneka kunye nokuchaneka kovavanyo.Ukuba kufuneka wenze umsebenzi wokulinganisa, nceda ugcine iirekhodi zeparameter yokuqala ukuze ubuyisele idatha yokuqala.

5. Ukulinganisa kweCapillary

Faka ityhubhu yokuvavanya engenanto kwiNombolo ye-1 ye-tray yesampula kwaye ungeze i-3ml yamanzi adibeneyo, cofa imenyu ye-"Settings", kwaye ukhethe.

"Ukulungiswa kweCapillary".Emva koko ucofe u-"Recalibrate" kwaye "Kulungile".Isixhobo siya kwenza ngokuzenzekelayo ukulinganisa okuthathu.Emva kokulinganisa, cofa u-"Yamkela", kwaye ekugqibeleni ucofe u-"Ewe" ukugcina iiparamitha zokulinganisa ezintsha.

  • malunga nathi01
  • malunga nathi02
Bhala umyalezo wakho apha kwaye uwuthumele kuthi

IINDIDI zeeMveliso

  • I-Semi Automated Blood Rheology Analyzer
  • Iikhithi zoLawulo lwe-Blood Rheology
  • I-Automated Blood Rheology Analyzer
  • I-Automated Blood Rheology Analyzer
  • I-Automated Blood Rheology Analyzer
  • I-Automated Blood Rheology Analyzer